Title: | Passive neutron area monitor with CR39 |
Authors: | Vega Carrillo, Héctor René Barquero, Raquel Mercado Sánchez, Gema Alejandrina |
Issue Date: | Jul-2013 |
Publisher: | Novin Medical Radiation Institute |
Abstract: | In high-intensity, mixed and pulsed neutron fields the use of spectrometers or area monitors with ac ve detectors is useless; in these condi ons neutron measuring devices must have a passive detector. Here a passive neutron area monitor with CR39 track detector was designed and the response was calculated. Materials and Methods: The response of a passive neutron area monitor with CR39 track detector has been calculated with the MCNPX code. To increase the detec on efficiency a 10B converter was included. The response was calculated using 47 monoenerge c neutron sources. Results: A passive neutron area monitor using a CR39 with 10B converter was designed where fluence and H*(10) responses were calculated. Conclusion: The shape of the responses is similar to responses reported for neutron monitors with ac ve detectors. |
URI: | http://localhost/xmlui/handle/20.500.11845/715 https://doi.org/10.48779/4tn4-1e02 |
ISSN: | 2322-3243 2345-4229 |
Other Identifiers: | info:eu-repo/semantics/publishedVersion |
Appears in Collections: | *Documentos Académicos*-- UA Ciencias Nucleares |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Passive neutron.pdf | 392,54 kB | Adobe PDF | View/Open |
This item is licensed under a Creative Commons License