Title: Neutron area monitor with passive detector
Authors: Valero Luna, Claudia
Guzmán García, Karen Arlete
Borja Hernández, Candy Gretel
Hernández Dávila, Víctor Martín
Vega Carrillo, Héctor René
Issue Date: Jun-2012
Publisher: Sociedad Mexicana de Física, A.C. Departamento de Física, 2o. Piso, Facultad de Ciencias, UN
Abstract: Se utilizó el método de Monte Carlo para calcular las respuestas de un monitor de neutrones pasivo. Para detectar los neutrones térmicos el monitor tiene una l´amina de oro que se encuentra en el centro de un cilindro de polietileno. Los neutrones que inciden son moderados por los núcleos de polietileno que llega a la l´amina de oro con la energía para inducir la activación a través de la 197Au reacción (n, °) 198Au. El 197Au decae emitiendo rayos gamma de 0.411 MeV con una vida media de 2.7 días. La actividad inducida se destina a medir con un espectrómetro de rayos gamma con un detector de centelleo de 3” £ 3” NaI (Tl) y la actividad de saturación se correlaciona con la dosis equivalente ambiental. La respuesta se calculó para 47 fuentes de neutrones monoenergéticos desde 1£10¡9 a 20 MeV. La respuesta a la fluencia se comparó con la respuesta del monitor de ´area LB 6411.
Description: Using Monte Carlo methods the responses of a passive neutron monitor area has been calculated. To detect thermal neutrons the monitor has a gold foil that is located at the center of a polyethylene cylinder. Impinging neutrons are moderated by polyethylene nuclei reaching the gold foil with the energy to induce activation through the reaction 197Au(n, °) 198Au. The 198Au decays emitting 0.411 MeV gamma rays with a half life of 2.7 days. The induced activity is intended to be measured with a gamma-ray spectrometer with a 3” £ 3” NaI(Tl) scintillator and the saturation activity is correlated to the ambient dose equivalent. The response was calculated for 47 monoenergetic neutron sources ranging from 1£10¡9 to 20 MeV. Calculated fluence response was compared with the response of neutron monitor area LB 6411.
URI: http://localhost/xmlui/handle/20.500.11845/737
https://doi.org/10.48779/1p7g-0016
ISSN: 0035-001X
Other Identifiers: info:eu-repo/semantics/publishedVersion
Appears in Collections:*Documentos Académicos*-- UA Ciencias Nucleares

Files in This Item:
File Description SizeFormat 
Neutron area monitor.pdf291,18 kBAdobe PDFThumbnail
View/Open


This item is licensed under a Creative Commons License Creative Commons